Z-1044-2026 Class II Ongoing
FDA device recall Z-1044-2026 was initiated by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING on Dec 10, 2025 and is designated Class II. Reason for recall: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details… The recall status is ongoing. Affected quantity: 7 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 21, 2026
- Initiation Date
- Dec 10, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product description
SMV DSX New Line, System, Tomography, Computed, Emission
Reason for recall
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Distribution pattern
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
Code information
UDI/DI Not applicable: All serial numbers in distribution