Z-1044-2026 Class II Ongoing

Recalled by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — Tirat Carmel

Recall Details

Product Type
Devices
Report Date
January 21, 2026
Initiation Date
December 10, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

SMV DSX New Line, System, Tomography, Computed, Emission

Reason for Recall

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Distribution Pattern

US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece

Code Information

UDI/DI Not applicable: All serial numbers in distribution