Z-1057-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
March 21, 2018
Initiation Date
November 29, 2017
Termination Date
April 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,227 in total

Product Description

NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE F, RIGHT¿ " These devices are indicated for patients with: - Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. - Previous tibial condyle or plateau fractures with loss of anatomy or function. - Varus or valgus deformities. - Revision of previous arthroplasty procedures. " These devices are indicated for cemented use only. " The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.

Reason for Recall

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

Code Information

00-5762-016-52 Knee LPS-FLEX GSF POR FEM, F-R