Z-1059-2022 Class II Ongoing

Recalled by bioMerieux, Inc. — Durham, NC

Recall Details

Product Type
Devices
Report Date
May 18, 2022
Initiation Date
March 29, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21,757 systems

Product Description

VITEK 2 Systems and VITEK 2 with MYLA.

Reason for Recall

Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.

Distribution Pattern

U.S. distribution of the FSCA to NC. OUS distribution of the FSCA to subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Serbia, Sweden, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, and United Kingdom.

Code Information

VITEK 2 - Versions 8.01, 9.01, 9.02 (with or without 9MR2), 9.03 (with or without #9MR2). VITEK 2 with MYLA - Versions 4.8 and 4.9.