Z-1060-2013 Class II Terminated

Recalled by Polymedco, Inc — Cortlandt Manor, NY

Recall Details

Product Type
Devices
Report Date
April 10, 2013
Initiation Date
February 12, 2013
Termination Date
February 10, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
229,600 test strips

Product Description

Test Strips packaged in the OC-Light Manual iFOBT Kit, Catalog No. FOBSTR. For the rapid, qualitative detection of human hemoglobin in feces.

Reason for Recall

The test strips in the affected product lots of the OC Light immunological fecal occult blood test are being recalled because they may not be stable through their labeled shelf-life. There may be a possibility of obtaining false negative results when using the affected lots.

Distribution Pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada and Guam.

Code Information

Kit Lot # P207830 - Kit Expiration Date 2/28/2013 - Test Strip Lot # 18005M; Kit Lot # P210191 - Kit Expiration Date 4/30/2013 - Test Strip Lot # 1X009M Kit Lot # P210192 - Kit Expiration Date 4/30/2013 - Test Strip Lot # 1X010M; Kit Lot # P217033 - Kit Expiration Date 5/31/2013 - Test Strip Lot # 1Y011M; Kit Lot # P217034 - Kit Expiration Date 5/31/2013 - Test Strip Lot # 1Y012M; Kit Lot # P217035 - Kit Expiration Date 6/30/2013 - Test Strip Lot # 1Z013M; Kit Lot # P222808 - Kit Expiration Date 6/30/2013 - Test Strip Lot # 1Z013M.