Z-1065-2019 Class II Terminated

Recalled by Accelerate Diagnostics Inc — Tucson, AZ

FDA device recall Z-1065-2019 was initiated by Accelerate Diagnostics Inc on Mar 13, 2017 and is designated Class II. Reason for recall: Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to… The recall status is terminated (terminated May 21, 2021). Affected quantity: 81.

Recall details

Product Type
Devices
Report Date
Apr 3, 2019
Initiation Date
Mar 13, 2017
Termination Date
May 21, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
81

Product description

Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.

Reason for recall

Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.

Distribution pattern

Worldwide Distribution - US Nationwide North America and Middle East US: TX, IL, TN, GA, CA, AR, MD, VA, Washington DC, IA, PA, NY, WA, and MO. OUS (Foreign): UK, Spain, Italy, and Germany

Code information

REF: 10401008. Serial Numbers: 1324, 1325, 1431, 1432, 1084, 1087, 1367, 1362,1363,1615, 1616, 1358, 1359, 1453, 1354, 1357, 1599, 1176, 1178, 1353, 1356, 1576, 1577, 1578, 1322, 1323, 1454, 1594, 1598, 1288, 1290, 1688, 1690, 1691, 1694, 1798, 1799, 1800, 1801, 1135, 1151, 1326, 1327, 1477, 1478, 1479, 1480, 1619, 1620, 1617, 1618, 1489, 1501, 1687, 1573, 1574, 1144, 1145, 1247, 1249 ,1264, 1277, 1265, 1277, 1278, 1381, 1150, 1286, 1148, 1149, 1254, 1260, 1397, 1542, 1458, 1517, 1231, 1238, 1282, 1284, 1389

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1065-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Accelerate Diagnostics Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1065-2019. Report a data issue: hello@regdatalab.com.