Z-1068-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 21, 2024
- Initiation Date
- December 19, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3613 Cases
Product Description
Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T
Reason for Recall
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
Distribution Pattern
US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
Code Information
REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281