Z-1068-2024 Class II Ongoing

Recalled by OriGen Biomedical, Inc. — Austin, TX

Recall Details

Product Type
Devices
Report Date
February 21, 2024
Initiation Date
December 19, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3613 Cases

Product Description

Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T

Reason for Recall

Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

Distribution Pattern

US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE

Code Information

REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281