Z-1070-2025 Class I Ongoing

Recalled by Spectranetics Corporation — Colorado Springs, CO

Recall Details

Product Type
Devices
Report Date
February 26, 2025
Initiation Date
January 10, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,939

Product Description

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062

Reason for Recall

Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.

Distribution Pattern

Distribution US nationwide and countries of: NL, DE, NL, AU, IT, CH, GB, and AT

Code Information

UDI-DI: REF/UDI-DI/Batch: 154150041/00863328000189/332905 - 349049, 154150042/328641, REF: 154090042/322101, 156135061/00863328000103/332824 - 336121, 206135061/00850003494043/331281 - 336152, 206135062/328643 - 345567, 156135062/271330 - 283066, 206080062/321083 - 349035, 156080062/302251 - 349034