Z-1074-2018 Class II Terminated
FDA device recall Z-1074-2018 was initiated by Smiths Medical ASD Inc. on December 4, 2017 and is designated Class II. Reason for recall: Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. … The recall status is terminated (terminated June 12, 2020). Affected quantity: 1,077 units total.
Recall Details
- Product Type
- Devices
- Report Date
- March 21, 2018
- Initiation Date
- December 4, 2017
- Termination Date
- June 12, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,077 units total
Product Description
Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)
Reason for Recall
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
Distribution Pattern
Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
Code Information
a. Item No. 21-4474-24 (Lot No. 87X040, 87X117); b. Item No. 21-4475-24 (Lot No. 87X044); c. Item No. 21-4477-24 (Lot No. 87X097); d. Item No. 21-8466-24 (Lot No. 87X171); e. Item No. 21-8468-24 (Lot No. 87X172, 87X244, 87X180); f. Item No. 21-8469-24 (Lot No. 87X151); g. Item No. 21-8470-24 (Lot No. 87X173, 87X254, 87X194, 87X198, 87X237)