Z-1074-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 18, 2022
- Initiation Date
- April 26, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product Description
4 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. M4 Ser CX 311-20 120-240V 50/60Hz, item # 807QS0020CX6.
Reason for Recall
Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of Australia, Canada.
Code Information
UDI-DI: +B2809807QS0020CX60L. Serial Number: QS11571PLNBCX0020