Z-1076-2024 Class II Ongoing

Recalled by Maquet Medical Systems USA — Wayne, NJ

FDA device recall Z-1076-2024 was initiated by Maquet Medical Systems USA on Jan 16, 2024 and is designated Class II. Reason for recall: The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection. The recall status is ongoing. Affected quantity: 16520 worldwide; 6657 US.

Recall details

Product Type
Devices
Report Date
Feb 21, 2024
Initiation Date
Jan 16, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16520 worldwide; 6657 US

Product description

Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5.0, Product Code/Part Number 70106.9077. Used with the CARDIOHELP heart-lung machine.

Reason for recall

The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.

Distribution pattern

Worldwide - US Nationwide distribution.

Code information

UDI-DI 04058863076355 Lots 3000301222, 3000319782, 3000327602 ***Added 3/21/24*** 3000333098, 3000352360, 3000357174

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1076-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1076-2024. Report a data issue: hello@regdatalab.com.