Z-1078-2026 Class II Ongoing

Recalled by Elekta, Inc. — Atlanta, GA

Recall Details

Product Type
Devices
Report Date
January 21, 2026
Initiation Date
December 10, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Reason for Recall

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Distribution Pattern

US Nationwide distribution in the state of TX.

Code Information