Z-1078-2026 Class II Ongoing
FDA device recall Z-1078-2026 was initiated by Elekta, Inc. on December 10, 2025 and is designated Class II. Reason for recall: Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial tr… The recall status is ongoing. Affected quantity: 1.
Recall Details
- Product Type
- Devices
- Report Date
- January 21, 2026
- Initiation Date
- December 10, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product Description
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Reason for Recall
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
Distribution Pattern
US Nationwide distribution in the state of TX.
Code Information