Z-1086-2014 Class II Terminated

Recalled by Stryker Spine — Allendale, NJ

Recall Details

Product Type
Devices
Report Date
March 5, 2014
Initiation Date
December 18, 2013
Termination Date
August 25, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 units

Product Description

Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.

Reason for Recall

Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.

Distribution Pattern

US Nationwide Distribution

Code Information

D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242