Z-1087-2019 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-1087-2019 was initiated by Zimmer Biomet, Inc. on March 1, 2019 and is designated Class II. Reason for recall: Two lengths of the 6.5mm RingLoc Hip System Self-Tapping Bone Screw may have the potential for the 30mm screws to be packaged as 20mm screws and vice versa. The recall status is terminated (terminated May 27, 2020). Affected quantity: 809 units total.

Recall Details

Product Type
Devices
Report Date
April 10, 2019
Initiation Date
March 1, 2019
Termination Date
May 27, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
809 units total

Product Description

Biomet RingLoc Hip System Self-Tapping Bone Screw 6.5mmx30mm-Low Profile Titanium, Accessory to total hip implant Item Number: 103533 Product Usage: Accessory to total hip implant

Reason for Recall

Two lengths of the 6.5mm RingLoc Hip System Self-Tapping Bone Screw may have the potential for the 30mm screws to be packaged as 20mm screws and vice versa.

Distribution Pattern

Worldwide Distribution - US Nationwide Foreign: CHILE CHINA COSTA RICA GUJRAT,INDIA JAPAN MEXICO CDMX SINGAPORE

Code Information

Lot Numbers/UDI Number 859960 (01) 00880304008700 (17) 280809 (10) 859960 859980 (01) 00880304008700 (17) 280805 (10) 859980 860120 (01) 00880304008700 (17) 280806 (10) 860120 860310 (01) 00880304008700 (17) 280823 (10) 860310 860580 (01) 00880304008700 (17) 280807 (10) 860580