Z-1091-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 25, 2022
- Initiation Date
- March 14, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15
Product Description
ATLAS Stim Headbox-NK For use only with the ATLAS Neurophysiology System. CAUTION--INVESTIGATIONAL DEVICE Limited by Federal (or United States) law to investigational use. Only for use in Non-Significant Risk studies under the supervision of an IRB.
Reason for Recall
Atlas Stim Headbox (ASHB) was delivered to US customers under a premise that the device had an Investigational Device Exemption (IDE) and was for research use only, but the manufacturer does not have an approved IDE by the FDA. The product has not undergone electrical safety testing and has not received regulatory clearance.
Distribution Pattern
U.S. Nationwide distribution in the states of AL, CA, IL, MN, NC, RI.
Code Information
ASHB Part No. 31-0601-0077 (SN0001, SN0002, SN0004, SN0005, SN0006, SN0007, SN0008, SN0009, SN0010, SN0012); ASHB Part No. 31-0601-0132(SN0003, SN0005, SN0006); ASHB Part No. 31-0601-0089(SN0004, SN0006);