Z-1122-2019 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-1122-2019 was initiated by Becton Dickinson & Company on March 7, 2019 and is designated Class II. Reason for recall: The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cann… The recall status is terminated (terminated May 12, 2020). Affected quantity: 6,509,600.

Recall Details

Product Type
Devices
Report Date
April 17, 2019
Initiation Date
March 7, 2019
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6,509,600

Product Description

BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010

Reason for Recall

The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.

Distribution Pattern

US distribution to the states of FL and NJ.

Code Information

Lot Numbers, UDI: 18041501, (17)210315(10)18041501, (01)50885403235915. 18042501, (17)210325(10)18042501, (01)50885403235915. 18050501, (17)210405(10)18050501, (01)50885403235915. 18051501, (17)210415(10)18051501, (01)50885403235915. 18052501, (17)210425(10)18052501, (01)50885403235915. 18060501, (17)210505(10)18060501, (01)50885403235915. 18061501, (17)210515(10)18061501, (01)50885403235915. 18062501, (17)210525(10)18062501, (01)50885403235915. 18070501, (17)210605(10)18070501, (01)50885403235915. 18072501, (17)210625(10)18072501, (01)50885403235915. 18080501, (17)210705(10)18080501, (01)50885403235915. 18081501, (17)210715(10)18081501, (01)50885403235915.