Z-1142-2024 Class II Ongoing

Recalled by Voco GmbH — Cuxhaven, N/A

FDA device recall Z-1142-2024 was initiated by Voco GmbH on Jan 25, 2024 and is designated Class II. Reason for recall: Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When… The recall status is ongoing. Affected quantity: 1530 units.

Recall details

Product Type
Devices
Report Date
Feb 28, 2024
Initiation Date
Jan 25, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1530 units

Product description

IonoStar Plus, Catalog (REF) Numbers: a) 2540, b)2543, c) 2544, d) 2545, e) 2546, f) 2547, g) 2549; dental cement

Reason for recall

Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.

Distribution pattern

Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.

Code information

a) 2540, UDI/DI +E22125401, Lot Numbers: 2338699, 2347154; b)2543, UDI/DI +E22125431, Lot Numbers: 2344046; c) 2544, UDI/DI +E22125441, Lot Numbers: 2342759, 2439537; d) 2545, UDI/DI +E22125451, Lot Numbers: 2348503; e) 2546, UDI/DI +E22125461, Lot Numbers: 2344045, 2349452; f) 2547, UDI/DI +E22125471, Lot Numbers: 2349453; g) 2549, UDI/DI +E22125491, Lot Numbers: 2342760, 2344016, 2344047, 2351233

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1142-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Voco GmbH? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1142-2024. Report a data issue: hello@regdatalab.com.