Z-1151-2014 Class II Terminated

Recalled by Megadyne Medical Products, Inc. — Draper, UT

FDA device recall Z-1151-2014 was initiated by Megadyne Medical Products, Inc. on Feb 27, 2014 and is designated Class II. Reason for recall: Megadyne Medical Products, Inc. is voluntarily recalling Hand Control 12Fr Suction Coagulator with 10ft cord and 132405 Holster, Sterile, Disposable, Lot 132405 due to the potential for a small number… The recall status is terminated (terminated Jul 9, 2014). Affected quantity: 494.

Recall details

Product Type
Devices
Report Date
Mar 19, 2014
Initiation Date
Feb 27, 2014
Termination Date
Jul 9, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
494

Product description

Hand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek¿ polyester peel pouch, 25 pouches per box. This device is intended to conduct monopolar electrosurgical energy from an electrosurgical generator (ESU) to target tissue during ENT and general surgical procedures. It is also a suction apparatus used to remove surgical smoke and fluids from the site. The device is intended for single use; it is not intended to be cleaned or reused.

Reason for recall

Megadyne Medical Products, Inc. is voluntarily recalling Hand Control 12Fr Suction Coagulator with 10ft cord and 132405 Holster, Sterile, Disposable, Lot 132405 due to the potential for a small number of these suction coagulators to self-activate immediately on connection to the electrosurgical generator and to remain activated until it is disconnected.

Distribution pattern

USA including FL, MN, LA, NJ, NM, PA, TX and Internationally to Canada.

Code information

Catalog Number: 0043-25 Lot Number: 132405 Exp. Date: 07-2018

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1151-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Megadyne Medical Products, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1151-2014. Report a data issue: hello@regdatalab.com.