Z-1155-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 1, 2017
- Initiation Date
- January 5, 2016
- Termination Date
- June 30, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 72 units in total
Product Description
Oxford Partial Knee System Phase 3 Tibial Template Right Medial Size A Product Usage: Instruments for use with the Oxford Uni Partial Knee System
Reason for Recall
Zimmer Biomet is conducting a medical device field action for various lot-specific Oxford Knee instruments due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
Distribution Pattern
Worldwide Distribution in the states of LA, NC, AR, MT, MO, CA, NY, KS, TX, MA, KY, OK, SC, NJ and the countries of Foreign: THAILAND, JAPAN, NETHERLANDS
Code Information
Item: 32-420826, Lots: ZB160801