Z-1166-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 1, 2013
- Initiation Date
- March 4, 2013
- Termination Date
- November 22, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25 units
Product Description
NeuViz Dual series CT Scanner System, Part number (PN): 989605651321. This is a computed X-ray, Tomography system. NeuViz Dual Multi-slice CT Scanner System is intended to produce cross-section images of head and body by computer reconstruction of X-ray transmission data taken at different angles.
Reason for Recall
User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.
Distribution Pattern
Nationwide distribution: USA including states of: California, Connecticut, North Carolina, Nebraska, Ohio, South Carolina, Texas, and Puerto Rico.
Code Information
Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.