Z-1191-2024 Class II Ongoing

Recalled by NxStage MDS Corporation — Lawrence, MA

FDA device recall Z-1191-2024 was initiated by NxStage MDS Corporation on Jan 29, 2024 and is designated Class II. Reason for recall: Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare… The recall status is ongoing. Affected quantity: 693.

Recall details

Product Type
Devices
Report Date
Mar 6, 2024
Initiation Date
Jan 29, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
693

Product description

NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.

Reason for recall

Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.

Distribution pattern

US domestic distribution to Dialysis Centers located in AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI. No international distribution.

Code information

UDI-DI M535CAR5350; Package DI M535CAR5355 Lots 20878001, 30978013, 30978032, 31178011, 31178019

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1191-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for NxStage MDS Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for NxStage MDS Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about NxStage MDS Corporation →

Source: openFDA record for Z-1191-2024. Report a data issue: hello@regdatalab.com.