Z-1194-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 4, 2015
- Initiation Date
- January 9, 2015
- Termination Date
- March 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product Description
NAMIC Convenience Kit, ANGIO DOUBLE SPIKE KIT, UPN H9656003028111, REF/Catalog No. 600302811, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136683. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
Reason for Recall
The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
Distribution Pattern
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
Code Information
Lot 4798703