Z-1199-2017 Class II Terminated
FDA device recall Z-1199-2017 was initiated by Acumed LLC on Jan 19, 2017 and is designated Class II. Reason for recall: The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V… The recall status is terminated (terminated May 8, 2017). Affected quantity: 15 units.
Recall details
- Product Type
- Devices
- Report Date
- Feb 22, 2017
- Initiation Date
- Jan 19, 2017
- Termination Date
- May 8, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1199-2017
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026