Z-1199-2017 Class II Terminated

Recalled by Acumed LLC — Hillsboro, OR

FDA device recall Z-1199-2017 was initiated by Acumed LLC on Jan 19, 2017 and is designated Class II. Reason for recall: The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V… The recall status is terminated (terminated May 8, 2017). Affected quantity: 15 units.

Recall details

Product Type
Devices
Report Date
Feb 22, 2017
Initiation Date
Jan 19, 2017
Termination Date
May 8, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product description

6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.

Reason for recall

The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135  Ti alloy) and was distributed.

Distribution pattern

Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain

Code information

Lot number: 383813

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1199-2017
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Acumed LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1199-2017. Report a data issue: hello@regdatalab.com.