Z-1208-2022 Class II Ongoing

Recalled by Bio-Rad Laboratories, Inc. — Hercules, CA

FDA device recall Z-1208-2022 was initiated by Bio-Rad Laboratories, Inc. on Apr 7, 2022 and is designated Class II. Reason for recall: The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM… The recall status is ongoing. Affected quantity: 1893.

Recall details

Product Type
Devices
Report Date
Jun 8, 2022
Initiation Date
Apr 7, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1893

Product description

Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000

Reason for recall

The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CD¿ROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall.

Distribution pattern

US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR OUS: Canada, Spain, France, Italy, India, Vietnam, Philippines

Code information

REF: 250-3020/Lot: AX90085, AY90085, AZ90085, BB90085, CB90085, DB90085; REF: 250-3000, UDI/DI: 00847817008276, Lot/Expiry: 64442741/ 2/28/2023, 64444208/ 2/28/2023, 64444218/ 3/26/2023, 64444715/ 3/31/2023, 64452929/ 3/31/2023, 64452922/ 5/31/2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1208-2022
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bio-Rad Laboratories, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1208-2022. Report a data issue: hello@regdatalab.com.