Z-1210-2025 Class II Ongoing

Recalled by Tyber Medical — Bethlehem, PA

Recall Details

Product Type
Devices
Report Date
March 5, 2025
Initiation Date
January 21, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
397

Product Description

A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, Right Model/Catalog Number: 770708042

Reason for Recall

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Distribution Pattern

Nationwide

Code Information

UDI-DI 00196449015535; Lot Numbers 267845, 261287, 265780, 24033BU01