Z-1255-2017 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

Recall Details

Product Type
Devices
Report Date
March 8, 2017
Initiation Date
January 10, 2017
Termination Date
July 16, 2019
Voluntary/Mandated
N/A
Product Quantity
1937 kits

Product Description

Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack) Cat No. 04800840, SMN 10285905 (6 pack - REF)

Reason for Recall

Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.

Distribution Pattern

Nationwide. Afghanistan, Albania, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Martinique, Mexico, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Vatican, Vietnam

Code Information

Lot Numbers: 07996A90 21172A90 27221A90 44366A90 63785A90 Exp. Date: 2017/07/24