Z-1256-2026 Class II Ongoing

Recalled by Vortex Surgical Inc. — Saint Charles, MO

FDA device recall Z-1256-2026 was initiated by Vortex Surgical Inc. on Dec 16, 2025 and is designated Class II. Reason for recall: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to… The recall status is ongoing. Affected quantity: 14,789 (8651 US; 6138 OUS).

Recall details

Product Type
Devices
Report Date
Feb 11, 2026
Initiation Date
Dec 16, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,789 (8651 US; 6138 OUS)

Product description

Vortex Surgical Lindsell Sutured IOL Marker, VS0390 - Scleral Marker

Reason for recall

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Distribution pattern

Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.

Code information

Pouch UDI 810123483655 Box UDI 810123483662 Lot 2408035

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1256-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1256-2026. Report a data issue: hello@regdatalab.com.