Z-1262-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 26, 2020
- Initiation Date
- January 15, 2020
- Termination Date
- March 2, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9,049,725
Product Description
i-STAT CHEM8+ cartridges (blue), List No. 09P31-25. UDI (01) 00054749001910 - Product Usage: The i-STAT CHEM8+ cartridge may be used for the quantitative measurement of sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine, and hematocrit in venous, arterial or capillary whole blood.
Reason for Recall
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the facilities that perform these types of tests. The i-STAT CHEM8+ cartridge contains nine measured assays (sodium, potassium, chloride, blood urea nitrogen (BUN), ionized calcium, TCO2, glucose, creatinine and hematocrit) and is categorized as a CLIA waived test for venous whole blood samples, meaning this cartridge could be used in a facility with a CLIA Certificate of Waiver as a waived test.
Distribution Pattern
US Nationwide distribution.
Code Information
All lots are affected