Z-1277-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 15, 2013
- Initiation Date
- November 10, 2009
- Termination Date
- July 5, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 608
Product Description
Olympus SurgMaster UES-40 electrosurgical unit ("UES-40") Product Usage: Designed for use in a medical facility under the supervision of a trained physician. It has been designed for general (open) and endoscopic surgery including urology, gynecology, respiratory and gastroenterology in conjunction with Olympus designed electrosurgical accessories, endoscopes (fiberscopes, videoscopes and rigid scopes) applicable for electrosurgery (cutting and coagulation), light sources and other ancillary equipment.
Reason for Recall
Olympus initiated a software upgrade and enhancement in 2009 and completed the action in 2010. The upgrade and enhancement were initiated due to complaints of overheating, smoke vapors and failure to operate associated with the Olympus SurgMaster electrosurgical unit, model UES-40 ("UES-40") when used in conjunction with the Olympus WA22557C electrode. The complaint investigations revealed that
Distribution Pattern
Nationwide Distribution - USA Nationwide including the states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, and WY.
Code Information
Model UES-40, all serial numbers prior to 7925334 that have not been upgraded.