Z-1285-2021 Class II
Recall Details
- Product Type
- Devices
- Report Date
- March 31, 2021
- Initiation Date
- February 26, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 71,473 units
Product Description
Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (SMALL) C3, REF: D138501, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016253 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.
Reason for Recall
There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism
Distribution Pattern
Worldwide distribution.
Code Information
All Lots