Z-1287-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 25, 2015
- Initiation Date
- February 4, 2015
- Termination Date
- October 30, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product Description
Signature TKA GDE/MDL Set 04-05 (Femur & Tibia) The Signature Patient-Specific Surgical Guides are intended to be used as part of the Signature Personalized Patient Care System. Lt is intended to be used as a guide during the surgical procedure for total and partial knee arthroplasty.
Reason for Recall
Images belonging to a different patient were used for the production of the patient specific surgical guide. The incorrect guide may cause a delay in surgery to accommodate the preparation and use of traditional instrumentation.
Distribution Pattern
Worldwide Distribution to the Netherlands only.
Code Information
Lot# 124471