Z-1321-2022 Class II Ongoing

Recalled by Radiometer Medical ApS — Bronshoj, N/A

Recall Details

Product Type
Devices
Report Date
July 13, 2022
Initiation Date
May 18, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1489 analyzers

Product Description

ABL800 Flex Analyzer Model Numbers 393-800 and 393-801

Reason for Recall

There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Bermuda, Canada, Italy, Norway, Austria, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, China, Croatia, Denmark, Iceland, Germany, Spain, Portugal, India, Brunei, Japan, United Kingdom, Ireland, Russia, Hungary, Australia, New Zealand, Tokelau, Switzerland, France, Czech Republic, Slovakia, South Africa, Botswana, Namibia, Poland, Turkey, Sweden, Finland, Estonia, Kazakhstan, Malaysia, Hong Kong, Mexico, South Korea, Singapore.

Code Information

All ABL800 Flex Analyzers containing software versions below V6.19 MR2; UDI/DI ABL810, 820, and 830, ABL805, ABL815, ABL825, ABL835: 05700693938004 ; UDI/DI ABL817, ABL827, ABL837: 05700693938011