Z-1322-2018 Class III Terminated

Recalled by Clinical Diagnostic Systems — Rochester, NY

Recall Details

Product Type
Devices
Report Date
April 18, 2018
Initiation Date
January 11, 2018
Termination Date
September 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)

Product Description

VITROS Immunodiagnostic Products TSH Reagent Pack

Reason for Recall

VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.

Distribution Pattern

Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.

Code Information

Lot 5470, expiry date: 24-May-2018 Lot 5430, expiry date: 01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017,