Z-1322-2018 Class III Terminated

Recalled by Clinical Diagnostic Systems — Rochester, NY

FDA device recall Z-1322-2018 was initiated by Clinical Diagnostic Systems on Jan 11, 2018 and is designated Class III. Reason for recall: VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency… The recall status is terminated (terminated Sep 8, 2020). Affected quantity: Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396).

Recall details

Product Type
Devices
Report Date
Apr 18, 2018
Initiation Date
Jan 11, 2018
Termination Date
Sep 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)

Product description

VITROS Immunodiagnostic Products TSH Reagent Pack

Reason for recall

VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.

Distribution pattern

Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.

Code information

Lot 5470, expiry date: 24-May-2018 Lot 5430, expiry date: 01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1322-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Clinical Diagnostic Systems? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1322-2018. Report a data issue: hello@regdatalab.com.