Z-1323-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 1, 2015
- Initiation Date
- February 20, 2015
- Termination Date
- February 26, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product Description
Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.
Reason for Recall
The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).
Distribution Pattern
US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.
Code Information
Serial Numbers: CA0N0000FT, CA0N0000H0, CA0N0000H7, CA0N0000J6, CA0N0000KJ, CA0N0000KK, CA0N0000KL, CA0N0000FW, CA0N0000G5, CA0N0000G6, CA0N0000H1, CA0N0000FR, CA0N0000FZ, CA0N0000KF, CA0N0000GH