Z-1323-2015 Class II Terminated

Recalled by ArthroCare Medical Corporation — Austin, TX

Recall Details

Product Type
Devices
Report Date
April 1, 2015
Initiation Date
February 20, 2015
Termination Date
February 26, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product Description

Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.

Reason for Recall

The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).

Distribution Pattern

US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.

Code Information

Serial Numbers: CA0N0000FT, CA0N0000H0, CA0N0000H7, CA0N0000J6, CA0N0000KJ, CA0N0000KK, CA0N0000KL, CA0N0000FW, CA0N0000G5, CA0N0000G6, CA0N0000H1, CA0N0000FR, CA0N0000FZ, CA0N0000KF, CA0N0000GH