Z-1329-2020 Class II Terminated

Recalled by Philips Medical Systems Gmbh, DMC — Hamburg

FDA device recall Z-1329-2020 was initiated by Philips Medical Systems Gmbh, DMC on Dec 19, 2019 and is designated Class II. Reason for recall: When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos… The recall status is terminated (terminated Oct 26, 2026). Affected quantity: 131.

Recall details

Product Type
Devices
Report Date
Feb 26, 2020
Initiation Date
Dec 19, 2019
Termination Date
Oct 26, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
131

Product description

CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.

Reason for recall

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Distribution pattern

Worldwide distribution  US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Code information

(domestic only) SN374 SN17081704 SN17121782 SN18011800 SN18081862 SN18061846 SN18111907

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1329-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Medical Systems Gmbh, DMC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Medical Systems Gmbh, DMC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Medical Systems Gmbh, DMC →

Source: openFDA record for Z-1329-2020. Report a data issue: hello@regdatalab.com.