Z-1331-2022 Class II Ongoing
FDA device recall Z-1331-2022 was initiated by Arcoma AB on May 5, 2021 and is designated Class II. Reason for recall: There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC). The recall status is ongoing. Affected quantity: 71 systems.
Recall details
- Product Type
- Devices
- Report Date
- Jul 13, 2022
- Initiation Date
- May 5, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 71 systems
Product description
Reason for recall
Distribution pattern
US Nationwide distribution in the states of NM, DE, TX, VT, FL, MO, UT, PA, OR, IN, RI, MI, IL, LA, OH, CO, SC, VA, WA, CA, AZ, NV, AR and the country of Canada.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1331-2022
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026