Z-1331-2022 Class II Ongoing

Recalled by Arcoma AB — Vaxjo, N/A

FDA device recall Z-1331-2022 was initiated by Arcoma AB on May 5, 2021 and is designated Class II. Reason for recall: There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC). The recall status is ongoing. Affected quantity: 71 systems.

Recall details

Product Type
Devices
Report Date
Jul 13, 2022
Initiation Date
May 5, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
71 systems

Product description

Omnera 400T Digital Radiographic System

Reason for recall

There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).

Distribution pattern

US Nationwide distribution in the states of NM, DE, TX, VT, FL, MO, UT, PA, OR, IN, RI, MI, IL, LA, OH, CO, SC, VA, WA, CA, AZ, NV, AR and the country of Canada.

Code information

UDI/DI:07350008750050; 2001-2003, 2006-2044, 2046-2056, 2118-2126, 2128-2131, 2134-2160, 2164-2175, 2177-2195, 2197-2208, 2210-2212, 2214-2224, 2226-2236.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1331-2022
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Arcoma AB? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1331-2022. Report a data issue: hello@regdatalab.com.