Z-1396-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 5, 2013
- Initiation Date
- November 10, 2012
- Termination Date
- October 29, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7
Product Description
Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Reamer, 6", Part Number 2110-0504 Product Usage: Facilitate placement of bone cement in weakened or diminished bone
Reason for Recall
There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.
Distribution Pattern
USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Code Information
Manufacturer Part Number 2110-0504, Lot Number E909007R