Z-1411-2021 Class II Ongoing

Recalled by Covidien — Ponce, PR

FDA device recall Z-1411-2021 was initiated by Covidien on March 5, 2021 and is designated Class II. Reason for recall: Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which co… The recall status is ongoing. Affected quantity: 59,828 devices.

Recall Details

Product Type
Devices
Report Date
April 21, 2021
Initiation Date
March 5, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
59,828 devices

Product Description

Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

Reason for Recall

Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.

Distribution Pattern

Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.

Code Information

UDI-Device Identifier (GTIN/UPN): 10884523000887, 20884523000884 UDI-Device Identifier (GTIN/UPN): 10884523000900, 20884523000907 UDI-Device Identifier (GTIN/UPN): 10884523000917, 20884523000914