Z-1461-2020 Class II Terminated

Recalled by Response Biomedical Corp. — Vancouver

FDA device recall Z-1461-2020 was initiated by Response Biomedical Corp. on Mar 15, 2019 and is designated Class II. Reason for recall: The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house… The recall status is terminated (terminated Jun 9, 2022). Affected quantity: 543 units.

Recall details

Product Type
Devices
Report Date
Mar 18, 2020
Initiation Date
Mar 15, 2019
Termination Date
Jun 9, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
543 units

Product description

Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP¿ platform.

Reason for recall

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.

Distribution pattern

Worldwide distribution: US Nationwide distributions in the states of IN, GA, OK, KS, AZ, SC, NE, NC, MI, FL, CA, MN, and AR. Countries of Canada, Italy, Malaysia, Kuwait, Saudi Arabia, UAE, Ukraine, Russia, Dominican Republic, France, Mexico, Colombia, Singapore, China, Serbia, Romania, and Philippine.

Code information

Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318). UDI Code: (01) 1 0627966 00508 4 (10) 070318 (17)300919

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1461-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.