Z-1462-2015 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 29, 2015
- Initiation Date
- March 25, 2015
- Termination Date
- March 17, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- ~4,154 units distributed in US - both part numbers
Product Description
TIGERPAW System II, Part number C-TP-1509 (9 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left Atrial Appendage (LAA).
Reason for Recall
Increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or may not be caused by incomplete closure of the TigerPaw System II fastener.
Distribution Pattern
Worldwide Distribution: US (nationwide) and countries of:Australia, Bahrain, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy Netherland, Norway, Poland, Qatar, Slovenia, Spain, Saudi Arabia, Sweden, Switzerland, Turkey and United Kingdom.
Code Information
Part number C-TP-1509 (9 connector) All serial numbers/lots