Z-1471-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 3, 2023
- Initiation Date
- April 5, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6803 systems
Product Description
Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower
Reason for Recall
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Distribution Pattern
Worldwide - US Nationwide distribution.
Code Information
UDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower