Z-1471-2023 Class II Ongoing

Recalled by Draeger Medical Systems, Inc. — Andover, MA

Recall Details

Product Type
Devices
Report Date
May 3, 2023
Initiation Date
April 5, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6803 systems

Product Description

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Reason for Recall

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

UDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower