Z-1476-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 10, 2024
- Initiation Date
- February 21, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7482 kits
Product Description
ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.
Reason for Recall
The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IL, KS, LA, MI, MO, MS, NE, NY, OH, OR, PA, TX, VA, and WI. The countries of Angola, Austria, Belarus, Belgium, Brazil, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Mali, Mexico, Morocco, Poland, Portugal, Peoples Republic of China, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Switzerland, Taiwan, Tajikistan, Tanzania, Trinidad & Tobago, Turkey, Uganda, Uruguay, Uzbekistan, and Vietnam.
Code Information
LN 2K43-25 - Lot number 50808UN23, exp. 11/30/2024, UDI (01)00380740003302 (17)241130(10)50808UN23; LN 2K43-20 - Lot number 60104UN23, exp. 11/30/2024, UDI (01)00380740003296 (17)241130(10)60104UN23.