Z-1476-2024 Class II Ongoing

Recalled by Abbott Laboratories — Abbott Park, IL

Recall Details

Product Type
Devices
Report Date
April 10, 2024
Initiation Date
February 21, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7482 kits

Product Description

ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.

Reason for Recall

The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IL, KS, LA, MI, MO, MS, NE, NY, OH, OR, PA, TX, VA, and WI. The countries of Angola, Austria, Belarus, Belgium, Brazil, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Mali, Mexico, Morocco, Poland, Portugal, Peoples Republic of China, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Switzerland, Taiwan, Tajikistan, Tanzania, Trinidad & Tobago, Turkey, Uganda, Uruguay, Uzbekistan, and Vietnam.

Code Information

LN 2K43-25 - Lot number 50808UN23, exp. 11/30/2024, UDI (01)00380740003302 (17)241130(10)50808UN23; LN 2K43-20 - Lot number 60104UN23, exp. 11/30/2024, UDI (01)00380740003296 (17)241130(10)60104UN23.