Z-1477-2026 Class II Ongoing

Recalled by Olympus Corporation of the Americas — Center Valley, PA

FDA device recall Z-1477-2026 was initiated by Olympus Corporation of the Americas on November 14, 2025 and is designated Class II. Reason for recall: Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator recepta… The recall status is ongoing. Affected quantity: 602 units.

Recall Details

Product Type
Devices
Report Date
March 11, 2026
Initiation Date
November 14, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
602 units

Product Description

Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Reason for Recall

Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.

Code Information

Model/Catalog Number: SPL-SR; UDI: 00821925043824; Serial Numbers: All;