Z-1480-2026 Class II Ongoing

Recalled by Staar Surgical AG — Nidau, N/A

Recall Details

Product Type
Devices
Report Date
March 11, 2026
Initiation Date
January 21, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7

Product Description

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Reason for Recall

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Distribution Pattern

International distribution to the countries of India, Iran, Korea, Saudi Arabia.

Code Information

UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988