Z-1508-2025 Class II Ongoing

Recalled by PYRAMES INC — Cupertino, CA

Recall Details

Product Type
Devices
Report Date
April 9, 2025
Initiation Date
February 24, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35

Product Description

Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007

Reason for Recall

Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.

Distribution Pattern

US distribution to states of: MA and CA

Code Information

Software version: 2.2.3 Part/UFI-DI/Lot: 54-0002/10860007282925/24020902, 24040302, 23103002; 51-0007/10860007282901/24011702, 24111301