Z-1534-2021 Class II Terminated

Recalled by Greiner Bio-One North America, Inc. — Monroe, NC

FDA device recall Z-1534-2021 was initiated by Greiner Bio-One North America, Inc. on March 29, 2021 and is designated Class II. Reason for recall: VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromb… The recall status is terminated (terminated October 10, 2023). Affected quantity: 1,049,700 units.

Recall Details

Product Type
Devices
Report Date
May 12, 2021
Initiation Date
March 29, 2021
Termination Date
October 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,049,700 units

Product Description

VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing.

Reason for Recall

VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromboplastin time, and fibrinogen assays.

Distribution Pattern

US Nationwide distribution in the states of KY, TX, LA, OH, AR, IL, SC, PA, FL, NC, AZ, CA, MI, NE, KS, MI, MO, NY, and NJ.

Code Information

Item Number: 454322; Lot # A21023B5; UDI: (Case) 39120017574485 and (Rack) 29120017574488; Expiration Date: 02/06/2022 Item Number: 454332; Lot # A210338Q; UDI: (Case) 39120017574607 and (Rack) 29120017574600; Expiration Date: 03/01/2022 Item Number: 454334; Lot # A210233V; UDI: (Case) 39120017574539 and (Rack) 29120017574532; Expiration Date: 02/01/2022 Item Number: 454334; Lot # A21023XW; UDI: (Case) 39120017574539 and (Rack) 29120017574532; Expiration Date: 02/13/2022