Z-1541-2023 Class II Ongoing

Recalled by Linkbio Corp. — Rockaway, NJ

Recall Details

Product Type
Devices
Report Date
May 10, 2023
Initiation Date
March 21, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

LINK SLED Knee System POREX ENDO SLED KNEE FEM COMP, MEDIUM Item Number: 15-2220/52

Reason for Recall

Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected

Distribution Pattern

NY, NV Foreign: Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland, Slovakia, Spain, United Kingdom, Germany, Austria, Argentina, Australia, China, India, Iraq, Jordan, Colombia, Mexico, Saudi Arabia, Ukraine, Vietnam, and Belarus

Code Information

UDI-DI: 04026575429097 All lot numbers within expiration date