Z-1548-2026 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 18, 2026
- Initiation Date
- February 6, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,407,465
Product Description
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i81-11BK, TRUE METRIX GO Kit (United Kingdom - mmol/L)/RF4i82-11BK, TRUE METRIX GO Starter Kit (United Kingdom - mmol/L)/RF4i82-12BK
Reason for Recall
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Distribution Pattern
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
Code Information
Device Item/UDI-DI: RF4001-01BK/311917178691, RF4007-01/50428560402, RF4019-01/708820721098, RF4023-01/21292009335, RF4031-01/41220136131, RF4048-01/75450843576, RF4066-01/11822002783, RF4209-01/36800234567, RF4H01-01BK/21292006112, RF4H01-40/21292009793. International: RF4i29-11BK/N/A, RF4i81-11BK/N/A, RF4i82-11BK/21292012229, RF4i82-12BK/21292009984. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX GO Owner's Booklet/ RF4TVH03 Rev 57 and prior, TRUE METRIX GO Owner's Booklet (English)/RF4TVH35 Rev 51 and prior, TRUE METRIX GO Owner's Booklet (Spanish)/RF4TVH35S Rev 51 and prior, TRUE METRIX GO Blood Glucose Monitoring System IFU Trividia House Brand/ RF4ITV16 Rev 55 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health UK Limited/ RF4UKT16 Rev 56 and prior, Trividia Health LATAM- TRUE METRIX GO System IFU/ RF4LAT16 Rev 51 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health Australia/ RF4AUT16 Rev 57 and prior,