Z-1552-2014 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 14, 2014
- Initiation Date
- April 14, 2014
- Termination Date
- July 16, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9,927 units
Product Description
TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-001 (inner tray). For use with the TX System Console as an ultrasonic surgical aspirator of soft tissue.
Reason for Recall
The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.
Distribution Pattern
Worldwide Distribution - USA (Nationwide) and Internationally to Panama and Singapore.
Code Information
Lot Numbers: 00614-03, 06513-02, 24913-01, 00713-06, 06514-02, 25313-01, 00814-02, 06914-01, 25513-02, 01114-02, 07213-04, 25613-01, 01514-03, 07713-18, 26113-06, 01714-04, 08113-04, 26313-04, 01813-02, 08413-13, 26713-04, 02314-08, 09313-05, 26813-01, 02413-04, 10213-07, 27013-10, 02714-02, 10613-01, 27613-02, 02914-01, 11213-07, 28013-05, 03014-02, 11613-06, 28313-02, 03113-02, 12113-02, 29513-03, 03114-02, 12713-03, 29713-01, 03414-03, 13413-01, 30913-07, 03514-05, 14413-05, 31113-08, 03714-03, 15013-04, 31513-02, 03814-06, 15413-02, 31713-05, 04114-02, 15413-03, 32213-01, 04213-01, 16513-04, 32413-01, 04214-05, 17013-02, 32613-01, 04414-03, 17013-09, 32913-07, 04514-02, 17613-08, 33613-09, 04814-07, 17913-16, 33813-05, 05014-02, 18413-05, 34013-03, 05113-01, 18413-06, 34313-01, 05113-02, 18613-03, 34413-04, 05114-09, 19213-03, 34513-04, 05514-01, 19713-04, 34713-08, 05614-04, 20713-06, 35113-01, 05713-03, 21813-01, 35213-03, 05714-10, 23913-01, 36013-01, 06313-05, 24113-04, 36113-02, 06314-04, 24213-04, and 36413-05.