Z-1552-2014 Class II Terminated

Recalled by Tenex Health Inc — Lake Forest, CA

FDA device recall Z-1552-2014 was initiated by Tenex Health Inc on Apr 14, 2014 and is designated Class II. Reason for recall: The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with… The recall status is terminated (terminated Jul 16, 2014). Affected quantity: 9,927 units.

Recall details

Product Type
Devices
Report Date
May 14, 2014
Initiation Date
Apr 14, 2014
Termination Date
Jul 16, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9,927 units

Product description

TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-001 (inner tray). For use with the TX System Console as an ultrasonic surgical aspirator of soft tissue.

Reason for recall

The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.

Distribution pattern

Worldwide Distribution - USA (Nationwide) and Internationally to Panama and Singapore.

Code information

Lot Numbers: 00614-03, 06513-02, 24913-01, 00713-06, 06514-02, 25313-01, 00814-02, 06914-01, 25513-02, 01114-02, 07213-04, 25613-01, 01514-03, 07713-18, 26113-06, 01714-04, 08113-04, 26313-04, 01813-02, 08413-13, 26713-04, 02314-08, 09313-05, 26813-01, 02413-04, 10213-07, 27013-10, 02714-02, 10613-01, 27613-02, 02914-01, 11213-07, 28013-05, 03014-02, 11613-06, 28313-02, 03113-02, 12113-02, 29513-03, 03114-02, 12713-03, 29713-01, 03414-03, 13413-01, 30913-07, 03514-05, 14413-05, 31113-08, 03714-03, 15013-04, 31513-02, 03814-06, 15413-02, 31713-05, 04114-02, 15413-03, 32213-01, 04213-01, 16513-04, 32413-01, 04214-05, 17013-02, 32613-01, 04414-03, 17013-09, 32913-07, 04514-02, 17613-08, 33613-09, 04814-07, 17913-16, 33813-05, 05014-02, 18413-05, 34013-03, 05113-01, 18413-06, 34313-01, 05113-02, 18613-03, 34413-04, 05114-09, 19213-03, 34513-04, 05514-01, 19713-04, 34713-08, 05614-04, 20713-06, 35113-01, 05713-03, 21813-01, 35213-03, 05714-10, 23913-01, 36013-01, 06313-05, 24113-04, 36113-02, 06314-04, 24213-04, and 36413-05.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1552-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tenex Health Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Tenex Health Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Tenex Health Inc →

Source: openFDA record for Z-1552-2014. Report a data issue: hello@regdatalab.com.