Z-1552-2020 Class II Terminated

Recalled by QUIDEL CARDIOVASCULAR INC — San Diego, CA

Recall Details

Product Type
Devices
Report Date
April 1, 2020
Initiation Date
April 9, 2019
Termination Date
January 14, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
123 kits

Product Description

Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com

Reason for Recall

Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Distribution Pattern

US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL

Code Information

Lot Number: C3470AN Part Number 94413