Z-1559-2013 Class II Terminated

Recalled by Hitachi Medical Systems America Inc — Twinsburg, OH

Recall Details

Product Type
Devices
Report Date
June 26, 2013
Initiation Date
January 31, 2013
Termination Date
September 20, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45 system units

Product Description

Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.

Reason for Recall

Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.

Distribution Pattern

Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.

Code Information

The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive